Reichert IOPac Standard Uživatelská příručka

Procházejte online nebo si stáhněte Uživatelská příručka pro Měření, testování a kontrola Reichert IOPac Standard. Reichert IOPac Standard User`s guide [en] [es] [fr] Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - Standard

StandardUser’s Guide®

Strany 2

10 16010-101 Rev. CIOPac Setup (Continued)Parts IdentificationParts:1. Charging Power Cord2. USB Cord3. Power Adapters4. IOPac Standard5. Graffit

Strany 3 - 16010-101 Rev. C 3

16010-101 Rev. C 11Instructions for UseTurning On and OffThe power button is located on the front of the IOPac. Press it once to turn the instrument

Strany 4 - Warnings & Cautions

12 16010-101 Rev. CInstructions for Use (Continued)Start Measuring Central Corneal Thickness (CCT) MeasurementPreparation of the PatientThe patient sh

Strany 5 - 16010-101 Rev. C 5

16010-101 Rev. C 13System Menu ScreenThe system menu screen is accessed by touching the “Reichert Pachymetry” menu at the top and left of the screen w

Strany 6 - Explanation of Symbols

14 16010-101 Rev. CSystem Menu Screen (Continued)About PachymeterThis screen shows the software and firmware versions, and the device serial number.Ho

Strany 7 - Introduction

16010-101 Rev. C 15 CleaningExternal CleaningWARNING: DO NOT IMMERSE THE CONNECTOR OR THE MAIN UNIT IN ANY LIQUID OR DAMAGE TO THE UNIT WILLOCCUR.Cle

Strany 8 - Introduction (Continued)

16 16010-101 Rev. CTroubleshootingFrequently Asked QuestionsQ: The unit will not power up.A: Always ensure the IOPac is charged sufficiently. USE ONL

Strany 9 - IOPac Setup

16010-101 Rev. C 17General SpecificationsThe IOPac displays thickness measurements to the nearest 1 µm, and the expected accuracy of readings is ±5 µm

Strany 10 - Parts Identification

18 16010-101 Rev. CGeneral Specifications (Continued)Acoustic Output PowersAcoustic Output MI ISPTA.3(mW/cm2)ISPPA.3(W/cm2)Global Maximum Value 0.155±

Strany 11 - Instructions for Use

16010-101 Rev. C 19General Specifications (Continued)Tissue Exposure To Ultrasound EnergyThe ultrasound energy emitted by the IOPac is of low intensit

Strany 12 - VOID THE IOPAC WARRANTY

2009 Reichert, Inc. Reichert is a registered trademark of Reichert, Inc.The information contained in this document was accurate at time of publication

Strany 13 - System Menu Screen

20 16010-101 Rev. CAppendix A – Installing IOPac Support on a Windows PC (HotSync)It is good practice to backup or “synchronize” files from the Pachym

Strany 14 - HotSync the IOPac

16010-101 Rev. C 21Appendix A (Continued)The Install Wizard encourages the user to perform a first time HotSync operation after the software has been

Strany 15 - Cleaning

22 16010-101 Rev. CAppendix A (Continued)You should also select Skip on the following screen, as there is no additional software needed or recommended

Strany 16 - Troubleshooting

16010-101 Rev. C 23Appendix B - Procedures for Reinstalling Missing IOPac Program FilesIOPac Pachymeters come from the factory with all necessary pach

Strany 17 - General Specifications

24 16010-101 Rev. CEnd User Software License AgreementCAREFULLY READ THE FOLLOWING TERMS AND CONDITIONS BEFORE USING THIS PRODUCT. IT CONTAINS SOFTWA

Strany 18 - Definitions

16010-101 Rev. C 25Reichert WarrantyThis product is warranted by Reichert, Inc. against defective material and workmanship under normal use for a peri

Strany 20

16010-101 Rev. C 27Notes

Strany 21 - Appendix A (Continued)

28 16010-101 Rev. CReichert, Inc.3362 Walden AveDepew, NY 14043USAToll Free 888-849-8955Phone: 716-686-4500Fax: 716-686-4545email: [email protected]

Strany 22

16010-101 Rev. C 3Table Of ContentsWarnings & Cautions...

Strany 23

4 16010-101 Rev. CWarnings & CautionsReichert, Inc. (Reichert) is not responsible for the safety and reliability of this instrument when:• Assembl

Strany 24 - 24 16010-101 Rev. C

16010-101 Rev. C 5CAUTION: THE IOPAC IS A PRECISION ULTRASOUND-MEASURING INSTRUMENT. PROTECT THE INSTRUMENTAGAINST DUST AND MOISTURE, AND AVOID PHYS

Strany 25 - Reichert Warranty

6 16010-101 Rev. CSymbolsExplanation of SymbolsType BF Patient-applied parts are insulated.Catalog NumberSerial NumberAttention, See Instructions for

Strany 26 - 26 16010-101 Rev. C

16010-101 Rev. C 7IntroductionCongratulations on your purchase of the IOPac Standard pachymeter.This User’s Guide is designed as a training and refere

Strany 27 - 16010-101 Rev. C 27

8 16010-101 Rev. CIntroduction (Continued)User InterfaceThe IOPac is a Palm Powered device. For documentation on the Palm interface, Graffiti text en

Strany 28

16010-101 Rev. C 9IOPac SetupGreat care has been taken to deliver your new IOPac Standard pachymeter safely to you. The container and packagingwas sp

Komentáře k této Příručce

Žádné komentáře