16010-101 Rev. C 13
System Menu Screen
The system menu screen is accessed by touching the “Reichert Pachymetry” menu at the top and left of the screen with
the stylus. When the “Reichert Pachymetry” menu is selected, the following options are shown. The details of each
menu selection are indicated below:
Check Calibration
About Pachymeter
Check Calibration
It is recommended that you check the calibration approximately every 14 days.
From the main screen, tap Reichert Pachymeter located on the upper left corner of the IOPac screen with the stylus.
Select: Check Calibration from the pull down menu.
Place a drop of water (Do not use ultrasonic gel) on the calibration
phantom, tap the Calibrate button, and then place the probe lightly on
the calibration phantom. A tone signals calibration is complete and the
results of the calibration are reported next to the “Phantom Standard”
Note: The probe must be perpendicular to the calibration phantom
therefore some manual adjustment of probe alignment may be
required for the IOPac to obtain a reading. If the measured
thickness is slightly greater or less than the allowable range, try
redoing the calibration while exerting less pressure onto the
phantom. If the measured thickness consistently measures
outside of the tolerance range, please contact an authorized
Reichert dealer.